USDA probe FAQ: technical questions about cold-treatment probes
Answers to the questions reefer technicians, depots and exporters ask us most often about USDA probes: how many are used, where they go, how to calibrate them and what to check when something goes wrong.
Installation and probe placement
How many USDA probes are used per reefer container?
Cold-treatment reefers have three or four USDA probe inputs, depending on the unit. The treatment protocol defines how many probes are used and where they go in the load. Carrier documentation, for example, describes three USDA probes connected to the DataCORDER.
Where are USDA probes placed in the load?
At the positions required by the treatment protocol. Each probe tip must be fully inserted in the fruit pulp: not in the air between fruit and not touching the box.
Why must the probe tip be inside the fruit?
The cold-treatment record has to show pulp temperature. A tip left in the air between fruit reads air temperature, not fruit temperature, and the record no longer represents the treatment.
What is the difference between a USDA probe and the supply or return air sensors?
Supply and return air sensors control the refrigeration unit. A USDA probe measures the temperature inside the fruit, which is the value used to document that a cold treatment was completed.
Ice-bath calibration
How are USDA probes calibrated before a cold treatment?
In a controlled ice-point bath: a dense mix of finely crushed ice and distilled water, verified near 0.00 °C with a calibrated reference thermometer. The controller then stores a correction (offset) for each probe. Always follow the reefer manufacturer’s procedure for your controller and software version.
Why does a USDA probe fail calibration?
The most common cause is a bath that is not at 0.00 °C: too much free water, too little ice or not enough stabilization time. The controller assumes the bath is at 0.00 °C, so a bath that is a few tenths of a degree off can reject a good probe. Check the bath with a reference thermometer before replacing the probe. On current software, Carrier and Thermo King accept offsets within ±0.3 °C.
Can a calibration be accepted and still be wrong?
Yes. If the bath is not at 0.00 °C, the bath error can be stored as part of the probe offset and then shift the readings during the whole treatment. Our ice-bath calibration technical note explains this with worked examples.
What reference thermometer is recommended for the ice bath?
For high-quality verification we recommend a resolution of 0.01 °C or better, an expanded uncertainty of about ±0.05 °C or better at 0 °C, and a valid traceable calibration. These are recommended capabilities, not USDA regulatory requirements.
What should be recorded during calibration?
Date and time, container and controller, the reference thermometer and its reading, the reading and offset of each probe, the accept or reject result and how the bath was prepared. With this record a later deviation can be traced to the probe, the bath or the controller.
Troubleshooting
Why does a USDA probe give unstable or missing readings?
Usually a damaged or crushed cable, or moisture in the connector. Inspect the cable along its length and the connector before each trip.
What happens if the controller is set to the wrong USDA sensor type?
Controllers that support more than one USDA sensor type must be set to the type installed. With the wrong setting the readings are not correct, even with a good probe and a good bath.
Product and ordering
Which reefer brands are your USDA probes made for?
Our USDA probes are used on Carrier Transicold, Thermo King and Star Cool reefers, and are designed for Daikin. Compatibility depends on model, controller, sensor characteristic and connector, and is confirmed case by case.
Which OEM part numbers do your USDA probes replace?
Our USDA probes are offered as compatible with, and a replacement for, Carrier 12-50089-02, 12-50089-02PK4, 12-50089-00 and 12-00789-20PK4, Thermo King 41-8188 and 41-3107, Star Cool 818555C and the Daikin NTC (Standard) type with a 3-pin receptacle. PT100 sensors are not interchangeable. See the OEM USDA sensor cross-reference.
What information do you need to quote USDA probes?
The reefer make and model, the controller, the connector, the quantity and the destination market. See the USDA sensor specifications for the standard configuration.
Have a question that is not here?
Send us the reefer model, controller and your question. We reply with a technical answer.
Reference price: USD 18 per probe (15 m cable).
USDA PROBE is an independent manufacturer and is not affiliated with or endorsed by the United States Department of Agriculture, Carrier Transicold, Thermo King, Maersk Container Industry (Star Cool) or Daikin. Brand names are used only to identify equipment platforms. Product suitability should be verified for each equipment model and treatment program.